Delivery outside GER for business customers only
Quality directly from the manufacturer
Langzeitstabile Temperaturfühler über Jahre
Advice directly from experts
 

Hat rail transmitter + Thermocouple type J M8 PVC

72,53 € *

Prices plus VAT plus shipping costs of 4,95 € to GER (incl. VAT)

Delivery time 5 Workdays to BE, AT, NL (GER -1 workdays)

Repeated order: check settings carefully and change if necessary.
1.Hutschienen-Transmitter
Ihr passender Hutschienen-Transmitter
Transmitter für Thermoelemente
2.Bitte konfigurieren Sie Ihr Einschraub-Thermoelement
E - Material Anschlussleitung

Bitte stellen Sie sicher, dass die technischen Daten der Anschlussleitung, insbesondere hinsichtlich Temperatur und IP-Schutzklasse, zu Ihrer Applikation passen. Detailierte Informationen finden Sie in unseren Datenblättern.
Temperaturbereiche unserer Leitungen:
PVC -30 °C bis +90 °C

(E8010) PVC/PVC (2 x 0,22 mm²)
F - Länge Leitung
keine Auswahl
(F010) 1 m
(F020) 2 m
(F030) 3 m
(F040) 4 m
(F050) 5 m
(F100) 10 m
(F150) 15 m
(F200) 20 m
G - Stecker

Bitte beachten Sie den spezifizierten Temperaturbereich des Stecker / der Aderendenhülsen
isolierte Aderendhülsen: bis 80 °C
Mini-TE-Stecker: bis 220 °C
TE-Stecker: bis 220 °C

(G01) isolierte Aderendhülsen
H - Knickschutz
(H0) Ohne (Standard)
3.Wie viele Transmitter und Fühler benötigen Sie?
Price calculation
Basic price
72,53 €
Ihr passender Hutschienen-Transmitter:  
E - Material Anschlussleitung:  
F - Länge Leitung:  
G - Stecker:  
H - Knickschutz:  
Menge:  
Configuration is incomplete - please check it!
  • 803151-3191
Our Payments
MasterCardPayPalInvoiceVISAPrepayment

Advantages

Questions about the article?
Our employee will be happy to help you:
Julian Mäntele
Tel.: +49 (0)7653 96 59 771
E-Mail: jmaentele@testo-sensor.de
Active cable thermocouples are equipped with a temperature transmitter in DIN rail form. They...more
Product information "Hat rail transmitter + Thermocouple type J M8 PVC"

Active cable thermocouples are equipped with a temperature transmitter in DIN rail form. They convert the thermoelectric voltage into a standardized signal (mA). This means that the temperature measurement signals can always be transmitted safely, easily and even over longer distances to your control or display. Screw-in thermocouples with PVC cable are the most cost-efficient solution to measure temperature in pipes or vessels. PVC cables can be used up to +90 °C.

Directly to the downloads

________________________________

The APAQ R130 TC for thermocouples is a DIN rail transmitter for installation on standard DIN rails according to DIN EN50022. The sophisticated product design leaves sufficient space for mounting. It is optimally designed for use in plant and machine construction and is characterized by high accuracy, reliability, long-term stability and its robust product design. The transmitter is extremely insensitive to external influences such as vibration and EMC interference. Installation and commissioning are particularly user-friendly. For example, parameterization can be carried out wirelessly, conveniently and easily via the cell phone app using NFC technology. The monitoring functions such as sensor break monitoring, sensor short-circuit and measuring range monitoring can also be activated via this.

Here you can find our downloads for Hat rail transmitter for thermocouples APAQ-R130

pdf-download Data sheet Hat rail transmitter for thermocouples APAQ-R130
pdf-download Instruction manual Hat rail transmitter for thermocouples APAQ-R130
pdf-download Declaration of Conformity Hat rail transmitter for thermocouples APAQ-R130
 

Hat rail transmitter - Special features

Inputs and outputs Input: thermocouples
Output: 4 to 20mA, temperature linear output signal
Accuracy & Long-term stability Accuracy: depending on thermocouple
Long-term stability: max drift of ±0.05°C or ±0.05% of span |year
Design Robust - vibration and shock resistant design
Suitable for hat rails according to DIN EN50022
Compact housing design
Facilitated mounting
Parametrization Configuration - wireless via NFC technology
Free app for Iphone, Android & Huawei
Parameterization templates for fast mass configuration
Simple configuration with "copy and paste" in no time at all
Alarm function configurable via app
Sensor break monitoring
Sensor short circuit
Measuring range monitoring
 

Hat rail transmitter - Input Thermocouple

Measuring element Material / Raw Material Norm Maximum configurable measuring range Min. Span Accuracy
Type B Pt30Rh-Pt6Rh IEC 60584 0 °C to +1820 °C
32 °F to +3308 °F
+700 °C | +1292 °F +100 °C to +400 °C: ±10 °C
>400 °C: ±2 °C / 0,2% 1}
Type E NiCr-CuNi IEC 60584 -270 °C to +1000 °C
-454 °F to +1832 °F
+50 °C | +122 °F ±1 °C| ±0,2 % 1}
Type J Fe-CuNi IEC 60584 -210 °C to +1200 °C
-346 °F to +2192 °F
+50 °C | +122 °F ±1 °C| ±0,2 % 1}
Type K NiCr-Ni IEC 60584 -270 °C to +1300 °C
-454 °F to +2372 °F
+50 °C | +122 °F ±1 °C| ±0,2 % 1}
Type N NiCrSi-NiSi IEC 60584 -100 °C to +1300 °C
-148 °F to +2372 °F
+100 °C | +212 °F ±1 °C| ±0,2 % 1}
Type N NiCrSi-NiSi IEC 60584 -270 °C to -100 °C
-418 °F to +148 °F
+100 °C | +212 °F ±2 °C 1}
Type R Pt13Rh-Pt IEC 60584 -50 °C to +1750 °C
-58 °F to +3182 °F
+300 °C | +572 °F ±2 °C| ±0,2 % 1}
Type S Pt10Rh-Pt IEC 60584 -50 °C to +1750 °C
-58 °F to +3182 °F
+300 °C | +572 °F ±2 °C| ±0,2 % 1}
Type T Cu-CuNi IEC 60584 -270 °C to +400 °C
-454 °F to +752 °F
+50 °C | +122 °F ±2 °C| ±0,2 % 1}
1} of span (cold junction compensation error is not included)
Input impedance >10 MΩ
Max. wire loop resistance 5 kΩ
Cold Junction Compensation Internal or external
 

Hat rail transmitter - General information about the input

Zero adjustment Within measuring range
 

Hat rail transmitter - Output

Output type analog, temperature linear for TC
Output signal (mA) 4 to 20
Parametrization / Scaling Configurable via NFC
Load 818 Ω at 24 VDC
Connection type 2-wire
 

Hat rail transmitter - Time response

Closing time / Update time (Inor) ~150 - 300
Heating period After approx. 20 min. the typical accuracy is reached
Signal attenuation / Ajustable output filtering (Inor) 0,4 to 26 adjustable via App
Measuring cycle < 1
 

Hat rail transmitter - Sensor monitoring & sensor error

Sensor break / Short circuit Upscale (≥21.0 mA) or Downscale (≤3.6 mA)
Sensor failure effects (Inor) according to NAMUR NE43
 

Hat rail transmitter - Accuracy and stability

Temperature influence
TC Type B, E, J, K, R, S, T see table below
TC Type N (-100...+1300 °C) ±0,01 % < 4000 Ω 2} < ±0,02 % of span per °C
TC Type N (-270...-100 °C) ±0,01 % of span per °C
Influence of the sensor cable
Thermocouples and Voltage Negligible
Further data
Supply voltage influence <±0.005 % of span per volt
Long-term drift ±0.05 % of span per year
 

Hat rail transmitter - Type

Dimensions See drawing
Material | Flammability V0/HB, RoHS compliant
Mounting 35 mm hat rail according to DIN 50022 / EN 60715
Connection Single wires, max. 1,5 mm², AWG 16
Weight 40 All dimensions in mm
 

Hat rail transmitter - General data

Isolation none
Supply Voltage (VDC) 8 to 32, polarity protected
 

Hat rail transmitter - Ambient conditions

Ambient Temperatur Storage: -40 °C to +85 °C | -40 °F to +185 °F
Operating: -40 °C to +85 °C | -40 °F to +185 °F
Humidity 0 to 98 (non-condensing)
Protection Housing IP20 | Anschlussklemmen IP00
Vibration according to IEC 60068-2-6, Test Fc, 10bis2000 Hz, 10 g
Shock according to IEC-60068-2-27, test Ea
Environmental influences according to IEC 60068-2-31:2008, Test Ec
 

Hat rail transmitter - EMC

Standard Directive: 2014/30/EU | Harmonized standards: EN 61326-1, EN 61326-2-3
Immunity performance ESD, radiated EMC fields, magnetic fields: Criterion A
Burst, conducted RF: Criteria A
Overvoltage: standard deviation 1% of span
 

Hat rail transmitter - Factory configuration (if not ordered otherwise)

Input Pt100, 3-wire, 0 °C to 100 °C
Output (mA) 4 to 20
Sensor control Upscale (≥21.0 mA)
 

Hat rail transmitter - Mounting

You can easily mount the APAQ R130 hat rail transmitter on 35mm hat rails according to DIN EN50022. The mounting is easy because you can fix the transmitter on the rail without any tools.
Mounting material for the installation of the transmitter is available as accessory.
Important: To prevent measuring errors, the connecting screws for fastening the connecting cable must be tightened firmly.
Mounting and dismounting of the transmitter
(1) Fix the upper part of the transmitter on the rail
(2) Then press the lower part of the transmitter onto the rail. The electrical connection is made according to the wiring diagram
(4) To remove the transmitter, use a screwdriver and bend the latch downwards
 

Hat rail transmitter - Configuration | Parametrization


Before making a configuration of APAQ C130TC you need to do following:
Make sure that you have a mobile device with NFC communication activated.
Download the app INOR Connect to your mobile device.
Required versions:
iOS: iOS 13 or later and Iphone 7 or later
Android: Android 4.4 or later
Configuration procedure:
Launch the app by clicking on the App icon or holding your mobile device against the transmitter on the part of the device where NFC is located (only possible with Android). Click on “Read Configuration” and hold your mobile device against the transmitter as explained in the first section.
In the app you can edit the following:
Sensor type and number of wire circuits
Measuring range
Upscale or Downscale
sensor control
TAG-number
Password settings
In the configuration window you can enter and change the parameters. The selected configuration is transferred to the transmitter by clicking the "Send to transmitter" button. After the transfer is completed, the transmitter uses the new parameters.

________________________________

Screw-in thermocouples with PVC cable are the most cost-efficient solution to measure the temperature in pipes or vessels. PVC cables can be used up to +90 °C.

Here you can find our downloads for the Thermocouple type J M8 with PVC cable

pdf-download Data sheet Thermocouple type J M8 with PVC cable
pdf-download Instruction manual Thermocouple type J M8 with PVC cable
pdf-download Operating instruction Thermocouple Type J M8 with PVC cable
pdf-download Declaration of Conformity Thermocouple type J M8 with PVC cable
 

Thermocouple M8 - General Information

Measuring range -30 °C to +90 °C
Perm. °C range cable -30 °C to +90 °C
Accuracy -40 °C to +375 °C: ±1,5 °C according to DIN IEC 60584 Class 1
Response time t63 / t99: information is available on request
Pull-out force ≥ 30 N
                                       

Thermocouple M8 - Supply and output

Measuring element Thermocouple Type J
Measuring point Measuring point isolated
Measurement signal Thermovoltage
 

Thermocouple M8 - Ambient conditions

Protection class IP54 according DIN 60529 (depending on cable)
Humidity and moisture condensation resistance according to application-specific qualification
 

Thermocouple M8 - Certificates and Standards

Standards DIN EN 61326-1:2013 | DIN EN IEC 63000:2019-05
Directive RoHS 2011/65/EU | 2014/30/EU
Certificates Certificate of suitability (on request)
 

Thermocouple M8 - Technical Drawing

E - Material connection cable
F - Length connection cable
H - Bend protection
G - Connector
All dimensions in mm
 

Thermocouple M8 - Screw-in thread

Thread
Material thread Stainless steel 1.4301 | SUS 304
Length (mm) 9
Thread M8
Wrench size 12
Protection sleeve
Material Stainless steel 1.4301 | SUS 304
Diameter (mm)1} 4,5
Mounting length (mm) 17
1} Tolerance ± 0,1 mm
 

Thermocouple M8 - Thermocouple cable

Type Color Material (Outside / Inside) Perm. range °C Ø Outside (mm) Cross section (mm²) Color strand Ω / m
Thermocouple cable Type J1} PVC / PVC -30 to +90 3,8 ± 0,2 0,22  bk, wt 2,502}
Length (m) 1, 2, 3, 4, 5, 10, 15, 20
Insulation resistance: ≥ 100 MOhm at min. 100 VDC | 1}Color according to IEC 584 | 2}per thermocouple | Cable available as accessory
 

Thermocouple M8 - Connector

Insulated end ferrules (50 mm)
Mini-TC connector Type J bk
TC connector Type J bk
 

Thermocouple M8 - Delivery

Assembly instructions per process connection
Delivery and Packaging Probe, seperatly packaged in PE bag
Matching accessories Heat-conducting paste
 

Thermocouple M8 - Important assembly advices

Measurement errors can occur due to heat dissipation to the environment. To keep these as small as possible, we recommend immersing the protection sleeve of your temperature probe as deeply as possible in the medium to be measured during installation. The optimum installation depth should be 10-15 times the Ø of the protection sleeve or, when using an immersion sleeve, the Ø of the immersion sleeve. When installing in pipelines whose Ø does not have a sufficiently deep installation depth, you should install the probe either at an angle or in a pipe elbow. Make sure that there is sufficient space for the probe to be removed. 1) Installation with sufficient installation depth 2) Installation at an angle with small pipe Ø 3) Not like this: Minimum installation depth not reached

Please lay the cable in such a way that no water can penetrate the probe and with reserve loop (4). This allows you to extend the probe without disconnecting the electrical connection.

 

Qualification measurement series to ensure the function of our temperature probes over their service life

Qualifizierungsmessreihe Robust even with temperature changes Temperature probes are exposed to harsh environmental conditions, such as changing temperatures and humidity, and should still work reliably in the field for 15 years or more. Checkmark
Qualifizierungsmessreihe Resistant to heat and humidity Extremely varying ambient conditions cause immense stress for the temperature probe and the design must be suitably robust and qualified.        
Qualifizierungsmessreihe Long-term stable in any weather Continuous exposure to sunlight, snow, fog, cold and ice shall not impair the function of the temperature probe and compliance with the specification.        
Qualifizierungsmessreihe Robust even under high vibration load Depending on the application, compressors and fans generate strong vibrations which are transmitted to components and cables. Therefore the used temperature probes have to be vibration resistant (20g). Checkmark
Qualifizierungsmessreihe Reliable in practice: high pull-out force The robust design ensures the necessary cable pull-out resistance, even under vibration and higher temperatures. Checkmark
Qualifizierungsmessreihe Efficient through short response time Depending on the application, short response times may be necessary to ensure efficient control of the system. Our sensor design is optimised for this requirement and we prove response times.        
Qualifizierungsmessreihe Corrosion-resistant even in salt spray Temperature probes are also used in coastal regions. For this reason, the temperature probes must also be able to withstand salty air for a lifespan of 15 years. Checkmark
Checkmark

Applicable for this probe type

 

Properties
Material cable: PVC
industry: Automotive, Cement and building materials industry, Glass industry, Heating technology, Machinery and apparatus engineering, Plastics industry, Power generation, Steel and iron industry, Waste recycling/incineration
Temperature range: -30 °C to +90 °C
Process connection: M8
Thermocouple: Type J Class 1
Output: 4-20 mA
Input: Thermocouple
Type: Hat rail transmitter
Related links to "Hat rail transmitter + Thermocouple type J M8 PVC"
Available downloads:
Read, write and discuss reviews...more
Customer evaluation for "Hat rail transmitter + Thermocouple type J M8 PVC"
Write an evaluation
Jede Verbraucherbewertung wird vor ihrer Veröffentlichung auf ihre Echtheit überprüft, sodass sichergestellt ist, dass Bewertungen nur von Verbrauchern stammen, die die bewerteten Produkte auch tatsächlich erworben/genutzt haben. Die Überprüfung geschieht durch manuelle Überprüfung in Form eines Abgleichs der Bewertung mit der Bestellhistorie des Warenwirtschaftssystems, um einen vorangegangenen Produkterwerb zur notwendigen Bedingung für die Veröffentlichung zu machen.
Please enter these characters in the following text field.
I have taken note of the Privacy Policy.

The fields marked with * are required.

Hutschienen-Transmitter für Thermoelemente APAQ-R130 Vorderansicht Hat rail transmitter for thermocouples APAQ-R130
58,90 € *
Order number: 809700-2101
Add to shopping cart
Hutschienen-Transmitter + Einschraubfühler M8 PVC - Vorderansicht Hat rail transmitter + Screw-in probe M8 PVC
70,68 € *
Order number: 803151-3091
Add to shopping cart
Hat rail transmitter + Thermocouple type J M4 PVC Hat rail transmitter + Thermocouple type J M4 PVC
73,36 € *
Order number: 803151-1191
Add to shopping cart
Hat rail transmitter + Thermocouple type J M6 PVC Hat rail transmitter + Thermocouple type J M6 PVC
72,17 € *
Order number: 803151-2191
Add to shopping cart
Hat rail transmitter + Thermocouple type K M8 PVC Hat rail transmitter + Thermocouple type K M8 PVC
72,53 € *
Order number: 803151-3291
Add to shopping cart